Fluke 1524-P4 Specifications
Input Channels 2
Resolution PRTs and thermistors: 0.001°
Thermocouples: 0.01°
Logging 25 readings with statistics15,000 time and date stamped
Sample Interval (Normal) 1 second (simultaneous)
Typical Sample Interval (Fast Mode) 0.3 seconds
Sensor Types PRTs, RTDs, thermistors, and thermocouples
Thermocouple Types C, E, J, K, L, M, N, T, U, B, R, S
Statistics Maximum, minimum, average, standard deviation
Trending Scale: ±18°F (10°C), ±1.8°F (1°C), ±0.18°F (0.1°C), ±0.018°F (0.01°C), 10 minutes of real-time data
Power Requirements 3 x AA alkaline batteries, 12 V DC universal power supply
Computer Interface RS-232, 9940 I/O ToolKit software included
Environmental Conditions for Best Accuracy 55.4 to 91.4°F (13 to 33°C)
Millivolt Range and Accuracy 10 to 75 mV, ± (0.005% + 5 µV)
Internal Reference Junction Compensation ±0.36°F (±0.2°C)
Resistance Range and Accuracy 0 to 400 Ω ± (0.004% + 0.002 Ω)
200 Ω to 50 kΩ ± (0.01% + 0.5 Ω)
50 to 500 kΩ ± (0.03 %)
Temperature Coefficient, Voltage ±(0.001%/°C + 1 mV/°C)
Temperature Coefficient, Resistance 0.0008%/°C + 0.0004 Ω (0 to 400 Ω)
0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ)
0.06%/°C + 0.1 Ω (50 to 500 kΩ)
Excitation Current, Resistance 1 mA (0 to 400 Ω)
10 µA (0 Ω to 50 kΩ)
2 µA (50 to 500 kΩ)
Dimensions 3.75 x 7.9 x 1.86" (96 x 200 x 47 mm)
Weight 1.45 lbs (0.65 kg)
Safety EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04