| Range | 999.9 A |
| Resolution | 0.1 A |
| Accuracy | 2% + 5 digits (10 to 100 Hz) 2.5% + 5 digits (100 to 500 Hz) |
| Crest Factor (50/60 Hz) | 2.5 at 600 A 3 at 500 A 1.42 at 999.9 A Add 2% for C.F. >2 |
| Flexible Current Probe AC Current | |
| Range | 999.9 A 2500 A |
| Resolution | 0.1 A (≤999.9 A) 1 A (≤2500 A) |
| Accuracy | 3% RD + 5 digits (10 to 500 Hz) |
| Crest Factor (50/60 Hz) | 2.5 at 1400 A 3 at 1100 A 1.42 at 2500 A Add 2% for C.F. >2 |
| DC Current | |
| Range | 999.9 A |
| Resolution | 0.1 A |
| Accuracy | 2% RD + 5 digits (when using the ZERO function to compensate for offsets) |
| AC Voltage | |
| Range | 600 V 1000 V |
| Resolution | 0.1 V (≤600 V) 1 V (≤1000 V) |
| Accuracy | 1% RD + 5 digits (20 to 500 Hz) |
| DC Voltage | |
| Range | 600 V 1500 V |
| Resolution | 0.1 V (≤600 V) 1 V (≤1500 V) |
| Accuracy | 1% RD + 5 digits |
| mV DC | |
| Range | 500 mV |
| Resolution | 0.1 mV |
| Accuracy | 1% RD + 5 digits |
| Jaw Amps Frequency | |
| Range | 5 to 500 Hz |
| Resolution | 0.1 Hz |
| Accuracy | 0.5% RD + 5 digits |
| Trigger Level | 5 to 10 Hz, ≥10 A 10 to 100 Hz, ≥5 A 100 to 500 Hz, ≥10 A |
| Flexible Current Probe Amps Frequency | |
| Range | 5 to 500 Hz |
| Resolution | 0.1 Hz |
| Accuracy | 0.5% RD + 5 digits |
| Trigger Level | 5 to 20 Hz, ≥25 A 20 to 100 Hz, ≥20 A 100 to 500 Hz, ≥25 A |
| Voltage Frequency | |
| Range | 5 to 500 Hz |
| Resolution | 0.1 Hz |
| Accuracy | 0.5% RD + 5 digits |
| Trigger Level | 5 to 20 Hz, ≥5 V 20 to 100 Hz, ≥5 V 100 to 500 Hz, ≥10 V |
| DC Power | |
| Range | 600 kVA (600 V DC range) 1500 kVA (1500 V DC range) |
| Resolution | 0.1 kVA 1 kVA |
| Accuracy | 2% RD + 2 KVA 2% RD + 20 kVA |
| Resistance | |
| Range | 600 Ω 6000 Ω 60 kΩ |
| Resolution | 0.1 Ω (≤600 Ω) 1 Ω (≤6000 Ω) 0.01 kΩ (≤60.00 kΩ) |
| Accuracy | 1% RD + 5 digits |
| Capacitance | |
| Range | 100.0 μF 1000 μF |
| Resolution | 0.1 μF (≤100.0 μF) 1 μF (≤1000 μF) |
| Accuracy | 1% RD + 5 digits |
| Inrush Trigger Level | 5 A |
| Environmental Specifications | |
| Temperature | Operating: 14 to 122°F (-10 to 50°C) Storage: -40 to 140°F (-40 to 60°C) |
| Operating Humidity | Non-condensing (<10°C) ≤90% RH at 50 to 86°F (10 to 30°C) ≤75% RH at 86 to 104°F (30 to 40°C) ≤45% RH at 104 to 122°F (40 to 50°C) |
| Altitude | Operating: 6562' (2000 m) Storage: 39370' (12000 m) |
| General Specifications | |
| Maximum Voltage (Between Terminal and Earth Ground) | 1000 V AC 1500 V DC |
| Batteries | 2 AA IEC LR6 alkaline |
| Display | Dual display with backlight |
| Automatic power off | 20 minutes |
| Mechanical Specifications | |
| Jaw Opening | 1.3 " (34 mm) |
| Flexible Current Probe Diameter | 0.29" (7.5 mm) |
| Flexible Current Probe Cable Length | 5.9' (1.8 m) |
| Dimensions | 11 x 3.3 x 1.9" (281 x 84 x 49 mm) |
| Weight | 1.1 lbs (520 g) |