| Vertical Deflection | |
| Number of Channels | 2 |
| Bandwidth | 60 Mhz |
| Rise Time | 5.8 ns |
| Channel Architecture | All inputs fully insulated from each other and from ground Inputs may be activated in any combination |
| Input Coupling | AC or DC, with ground level indicator |
| Input Sensitivity | With 10:1 probe, 20 mV to 1000 V/div With 100:1 probe, 200 mV to 10 kV/div Direct (1:1), 2 mV to 100 V/div |
| Bandwidth Limiter | 20 MHz and 10 kHz |
| Polarity | Normal, inverted, variable |
| Vertical Resolution | 8 bit |
| Accuracy at 4 s to 10 μs/div | 5 mV/div to 100 V/div, ±(1.5% + 6 counts) 2 mV/div, ±(1.5 % + 10 counts) |
| Input impedance | 1 MΩ (±1 %) // 15 pF (±2.25 pF) |
| Input Voltage Range | |
| Rated Maximum Floating Voltage | CAT III 1000 V / CAT IV 600 V (maximum voltage between any contact and earth-ground voltage level) |
| Probe Input Voltage VPS410-II | CAT III 1000 V / CAT IV 600 V (maximum voltage between standard 10:1 probe tip and reference lead) |
| Probe Input Voltage VPS421 | CAT III 1000 V / CAT IV 600 V (maximum voltage between probe tip or reference lead to GND, 2000 V max between probe tip and reference lead) |
| Maximum BNC Input Voltage | CAT IV 300 V (maximum voltage on BNC input directly) |
| Memory Save and Recall | |
| Memory Locations (Internal) | 30 waveform memories plus 10 recording memories plus 9 screen copy memories |
| 30 Waveform Memories | Each memory can contain up to 2 or 4 waveforms plus corresponding setups |
| 10 Recording Memories | Each may contain: a 100 Screen Replay sequence, or a ScopeRecord Roll-mode recording (2 or 4 traces), or a TrendPlot recording of up to 4 measurements |
| External Data Storage | On PC, using FlukeView™-2 Software, or direct storage on external flash memory drive (maximum 32 GB) through USB host port |
| Screencopies | On PC, using FlukeView™-2 Software, or internally (in instrument) which can be copied on to external flash memory drive as .BMP-file, through USB host port |
| Volatility | Saving is done in non-volatile Flash-ROM and all data is secured, independent of battery or power status |
| Real-Time Clock | Provides date and time stamp information for ScopeRecord, for 100 Screen Replay sequences and for TrendPlot recordings |
| Case | |
| Design | Rugged, shock-proof with integrated protective holster. Handstrap and hangstrap included as standard Kensington lock supported to lock down instrument when left unattended |
| Drip and Dust Proof | IP 51 according to IEC60529 |
| Shock and Vibration | Shock 30 g, vibration (sinusoidal) 3 g / 0.03 g2/Hz (Random), according to MIL-PRF-28800F Class 2 |
| Display Size | 5.3 x 3.5" (133 x 90 mm) LCD |
| Resolution | 1120 pixels x 765 pixels |
| Brightness | User-adjustable, up to 300 cd/m2 |
| General Specifications | |
| Temperature | Operating: battery discharging: 32 to 104°F (0 to 40°C); battery charging: 0 to 40°C (32 to 104°F) Storage: -4 to 140°F (-20 to 60°C) |
| Humidity | 32 to 50°F (0 to 10°C): noncondensing 50 to 86°F (10 to 30°C): 95% (±5%) 86 to 104°F (30 to 40°C): 75% (±5%) 104 to 122°F (40 to 50°C): 45% (±5%) |
| Maximum Operating Altitude | CAT IV 600 V, CAT III 1000 V: up to 6600’ (2000 m) CAT IV 300 V, CAT III 600 V, CAT II 1000 V: up to 13000’ (4000 m) |
| Maximum Storage Altitude | 40,000’ (12 km) |
| Interfaces | Two USB-ports provided Ports are fully insulated from instrument’s floating measurement circuitry USB-host port directly connects to external flash memory drive (up to 32 GB) for storage of waveform data, measurement results, instrument settings and screen copies Alternatively, this USB-A port may be used to connect a WiFi Adapter for wireless PC connectivity A mini-USB-B is provided which allows for interconnection to PC for remote control and data transfer under PC-control using FlukeView-2. |
| Probe Calibration Output | Dedicated probe-cal output with reference contact provided, fully insulated from any measurement input channel Generator Output: 1.225 Vpp / 500 Hz square wave |
| Line Power | Universal mains adapter/battery charger BC190/830 included, with detachable 2-wire power cords 100 VAC to 240 VAC, ±10%, 50 to 60 Hz |
| Battery | Power: re-chargeable Li-Ion battery; battery swappable through easily accessible battery door at the rear of the instrument Type: BP291: 10.8V, 5000 mAh Charge indicator: battery has built-in status indicator for use with external charger, next to battery status indicator on instrument screen Operating time (with backlight low): Up to 7 hours using BP291 (included) Charging time: five hours for BP291 Power saving functions: Auto ‘power down’ with adjustable power down time; automatic ‘display off’ with adjustable power down time; on-screen battery power indicator |
| Dimensions | 10.5 x 7.6 x 2.8" (265 x 192 x 70 mm) |
| Weight | 4.8 lbs (2.2 kg) |


This website uses cookies (Learn more) and has a privacy policy (Learn more).