| Key Specifications | |
| Inputs | 3 Voltage + 3 Current |
| Sample Rate | 200 ks/s |
| Measurement Update Rate | 100 ms, 200 ms, 500 ms, 1s |
| Dual Analyzer Synchronization Mode | Able to extend to 6, 7 or 8 channels (using multiple instruments) |
| Data Trend Storage Rate | As per display rate |
| Main Function | Meter, Scope, Harmonic, Phasor, Trend |
| Measuring Parameters | RMS, DC Component, AC Component, Rectified Mean, Peak Value, Peak-Peak, Crest Factor, Form Factor, Fundamental Component, Fundamental Content, Harmonic Distortion, Harmonic Content, Harmonic Factor for voltages and currents, Active Power, Reactive Power, Apparent Power, Power Factor, Phase Shift, Efficiency, Impedance, Electric Energy, Charge/Discharge capacity (Ah), Frequency |
| Electrical | |
| Range | 10, 100, 1000 V |
| Crest Factor | CF ≤2 |
| Maximum Voltage | 10% overload |
| Input Impedance | 2 MΩ/10 pF (typical) |
| Temperature Coefficient | 0.05 (Spec)/k |
| Bandwidth | 1000 V range: 500 kHz 100 V range: 200 kHz 10 V range:100 kHz |
| CMRR | 120 dB at 50/60 Hz |
| Accuracy (% Reading + % Range) | |
| Range - 1000 V | DC: 0.1 + 0.1 AC (10 Hz to 1 kHz): 0.1 + 0.1 AC (10 kHz): 5 + 0.5 |
| Range - 100 V | DC: 0.1 + 0.1 AC (10 Hz to 1 kHz): 0.1 + 0.1 AC (10 kHz): 5 + 0.5 |
| Range - 10 V | DC: 0.1 + 0.2 AC (10 Hz to 1 kHz): 0.1 + 0.2 AC (10 kHz): 5 + 0.5 |
| Current | Crest factor: CF ≤2 Temperature coefficient: 0.05 x (Spec)/k Overload capacity: 10% overload CMRR: 120 dB at 50/60 Hz |
| Shunt (Current Input) | Measuring range: 0.1, 1, 10 A Input impedance: 0.025 Ω (typical) Bandwidth: 10 A - 500 kHz; 1 A - 200 kHz; 0.1 A - 100 kHz |
| BNC (Voltage Input) | Range: 0.1, 1, 10 V Input impedance: 100 kΩ/100 pF (typical) Bandwidth: 10 V - 500 kHz; 1 V - 200 kHz; 0.1 V - 100 kHz |
| Range - 10 A | DC: 0.1 + 0.2 AC (10 Hz to 1 kHz): 0.1 + 0.1 AC (10 kHz): 5 + 1 |
| Range - 1 A | DC: 0.1 + 0.5 AC (10 Hz to 1 kHz): 0.1 + 0.2 AC (10 kHz): 5 + 1 |
| Range - 0.1 A | DC: 0.1 + 2 AC (10 Hz to 1 kHz): 0.1 + 1 AC (10 kHz): 5 + 1 |
| Range - 10 V | DC: 0.1 + 0.1 AC (10 Hz to 1 kHz): 0.1 + 0.1 AC (10 kHz): 5 + 1 |
| Range - 1 V | DC: 0.1 + 0.2 AC (10 Hz to 1 kHz): 0.1 + 0.1 AC (10 kHz): 5 + 1 |
| Range - 0.1 V | DC: 0.1 + 1 AC (10 Hz to 1 kHz): 0.1 + 0.5 AC (10 kHz): 5 + 1 |
| General Specifications | |
| Operating Temperature | 14 to 122°F (-10 to 50°C) |
| Storage Temperature | -22 to 140°F (-30 to 60°C) |
| Operating Humidity | Non-condensing (<50°F [10°C]) ≤90% RH (at 50 to 86°F [10 to 30°C]) ≤75% RH (at 86 to 104°F [30 to 40°C]) ≤45% RH (at 104 to 122°F [40 to 50°C]) |
| Altitude | Operating: 6561.68' (2000 m) Storage: 39,370.1' (12,000 m) |
| Display | 5.7", TFT LCD, 640 x 480 |
| Battery | BP 291, 10.8 V/5000 mAh, 54Wh IEC 62133, UN38.3 Operating time: 10 Hours (on battery) |
| Communication Interface | USB/RS485 |
| Storage Capacity | 32GB |
| PC Software | Fluke Power Analyzer software |
| Ingress Protection | IP 50 (Terminals mated) according to IEC 60529 |
| Safety | IEC 61010-1: Pollution Degree 2 IEC 61010-2-030: CAT IV 600V, CAT III 1000 V |
| EMC | IEC 61326-1: Industrial IEC 61326-2-2 |
| Dimensions | 11.73 x 8.46 x 3.78" (298 x 215 x 96 mm) |
| Weight | 7.7 lbs (3.5 kg) |


This website uses cookies (Learn more) and has a privacy policy (Learn more).