| DC millivolts | |
| Range | 600.0 mV |
| Resolution | 0.1 mV |
| Accuracy | ±0.5% of reading + 2 counts |
| DC Volts | |
| Range | 6.000 V 60.00 V 600.0 V |
| Resolution | 0.001 V 0.01 V 0.1 V |
| Accuracy | ±0.5% of reading + 2 counts |
| Auto Volts | |
| Range | 600 mV |
| Resolution | 0.1 mV |
| Accuracy | 2% + 3 (dc. 45 to 500 Hz) 4% + 3 (500 Hz to 1 kHz) |
| AC millivolts True RMS | |
| Range | 600.0 mV |
| Resolution | 0.1 mV |
| Accuracy | 45 to 500 Hz: ±1.0% of reading + 3 counts 500 Hz to 1 kHz: ±2.0% of reading + 3 counts |
| AC Volts True RMS | |
| Range | 6.000 V 60.00 V 600.0 V |
| Resolution | 0.001 V 0.01 V 0.1 V |
| Accuracy | 45 to 500 Hz: ±1.0% of reading + 3 counts 500 Hz to 1 kHz: ±2.0% of reading + 3 counts |
| Continuity | |
| Range | 600 Ω |
| Resolution | 1 Ω |
| Accuracy | Beeper on <20 Ω off> 250 Ω; detects opens or shorts of 500 µs or longer |
| Ohms | |
| Range | 600 Ω 6 kΩ 60 kΩ 600 kΩ 6 MΩ 40 MΩ |
| Resolution | 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
| Accuracy | 0.9% + 1 |
| Diode Test | |
| Range | 2 V |
| Resolution | 0.001 V |
| Accuracy | 0.9% + 2 |
| Capacitance | |
| Range | 1000 nF 10 µF 100 µF 9999 µF 100 to 1000 µF >1000 µF |
| Resolution | 1 nF 0.01 µF 0.1 µF 1 µF |
| Accuracy | 1.9% + 2 1.9% + 2 5% + 20 |
| Lo-Z Capacitance | |
| Range | 1 nF to 500 µF |
| Accuracy | 10% + 2 typical |
| AC Amps True RMS (45 to 500 Hz) | |
| Range | 6.000 A 10.00 A 20 A overload for 30 seconds maximum |
| Resolution | 0.001 A 0.01 A |
| Accuracy | ±1.5% of reading + 3 counts |
| AC µAmps True RMS (45 Hz to 1 kHz) | |
| Range | 600.0 µA |
| Resolution | 0.1 µA |
| Accuracy | ±1.5% of readings + 3 counts (±2.5% of readings + 3 counts >500 Hz) |
| DC Amps | |
| Range | 6.000 A 10.00 A 20 A overload for 30 seconds maximum |
| Resolution | 0.001 A 0.01 A |
| Accuracy | ±1.0% of readings + 3 counts |
| Hz (V or A input) | |
| Range | 99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz |
| Resolution | 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz |
| Accuracy | ±0.1% of readings + 2 counts |
| General Specifications | |
| Maximum Voltage Between Any Terminal and Earth Ground | 600 V |
| Surge Protection | 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 |
| Fuse for A Input | 11 A, 1000 V FAST Fuse (Fluke PN 803293) |
| Display | Digital: 6000 counts, updates 4/sec |
| Bar Graph | 33 segments, updates 32/sec |
| Operating Temperature | 14 to 122°F (-10 to 50°C) |
| Storage Temperature | -40 to 140°F (-40 to 60°C) |
| Battery | 9 V alkaline, NEDA 1604A/IEC 6LR61 |
| Battery Life | 400 hours typical, without backlight |