| Measurement Accuracy | |
| Voltage DC | 30 V ; 0.01% + 2 mV (upper display) 20 V ; 0.01% + 2 mV (lower display) 90 mV ; 0.01% + 10 µV -10 to 75 mV ; 0.025% + 1 count (via TC connector) |
| Current DC | 24 mA ; 0.01%+ 2 mA |
| Resistance | 0 to 400 Ω ; ±0.015% |
| Frequency | 2 to 1000 CPM ; 0.05% + 0.1 cpm 1 to 1000 Hz ; 0.05% + 0.1 Hz 1 to 15 kHz ; 0.05% + 0.01 kHz Sensitivity ; 1 V peak-to-peak-minimum |
| Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
| Source Accuracy | |
| Voltage DC | 100 mV ; 0.01% + 10 µV 20 V ; 0.01% + 2 mV -10 to 75 mV ; 0.025% + 1 count (via TC connector) |
| Current DC | 24 mA (Source) ; 0.01% + 2 µA 24 mA (Simulate) ; 0.01% + 2 µA |
| Resistance | Range, accuracy (percent of reading + floor) and excitation current: 5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA 5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA 400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA 1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
| Frequency | 2 to 1000 CPM ; 0.05% 1 to 1000 Hz ; 0.05% 1 to 10 kHz ; 0.25% 10 to 15 kHz ; 0.5% Waveform ; 5 V p-p squarewave, -0.1 V offset |