| Pressure | |
| Range | -12 to 300 psi / -0.8 to 20 bar |
| Resolution | 0.01 psi, 0.001 bar |
| Accuracy | ±0.025% of full scale, 6 months ± 0.035% of full scale, 1 year dry air and non-corrosive gas only |
| Temperature Effect | Add ± 0.002% F.S./°C for temps outside of 59 to 95°F (15 to 35°C) No effect on accuracy on all functions from 15 to 35°C |
| mA DC (Simulation and Measurement) | |
| Range | 0 to 24 mA DC |
| Resolution | 0.001 mA |
| Accuracy | ± 0.015% of reading ± 2 counts |
| Max Load in mA Source | 1000 Ω |
| Max Voltage in mA Simulate | 26 V DC |
| Volts DC (Measurement only) | |
| Range | 30 V DC |
| Resolution | 0.001 V |
| Accuracy | ± 0.015% of reading ± 2 counts |
| Temperature (Measurement only, Pt100 RTD) | |
| Range | -50 to 150 °C (-50 to 302°F) |
| Resolution | 0.01 to 0.01°F |
| Accuracy | ± 0.1°C (0.2°F) |
| Loop Power Supply | |
| Range | 24 V DC |
| Accuracy | 24 mA at 24 V |
| Temperature Effect (all functions) | |
| Accuracy | Add ± 0.002% F.S./°C for temps outside of 15 to 35°C No effect on accuracy on all functions from 15 to 35°C |
| Connections | |
| Pressure | 1/8" female NPT |
| Electrical | Standard banana jacks |
| RTD | 4 pin miniature circular |
| External Module | 6 pin miniature circular |
| Power Requirements | |
| Battery | 8 x standard Alkaline AA batteries |
| Battery Life | 300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 psi; 100 pump cycles to 300 psi minimum |
| Physical | |
| Dimensions | 9.5 x 4.4 x 2.8" (24.13 x 11.18 x 7 cm) |
| Weight | With holster 2.5 lb (1.2 kg) |


This website uses cookies (Learn more) and has a privacy policy (Learn more).