| Range1 | Resolution | Accuracy2,3,4 | |
|---|---|---|---|
| 45 Hz to 500 Hz | 500 Hz to 1 kHz | ||
| 6.000 V | 0.001 V | 1.0 % + 3 | 2.0 % + 3 |
| 60.00 V | 0.01 V | ||
| 600.0 V | 0.1 V | ||
| 1000 V | 1 V | ||
| LCD w/backlight | 3.5 digits, 6000 counts, updates 4/sec |
|---|---|
| Battery type | 2 AA, NEDA 15 A, IEC LR6 |
| Battery life | 400 Hours |
| Memory | Record up to 65000 readings |
| RF communications | 2.4 GHZ ISM Band |
| RF communication range | Open air, unobstructed: Up to 20 m Obstructed, sheetrock wall: Up to 6.5 m Obstructed, concrete wall or steel electrical enclosure: Up to 3.5 m |
| Operating temperature | -10 °C to +50 °C |
| Storage temperature | -40 °C to +60 °C |
| Temperature coefficient | 0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C) |
| Operating humidity | 90 % at 35 °C, 45 % at 40 °C, 45 % at 50 °C |
| Altitude | Operating: 2,000 m Storage: 12,000 m |
| EMC | EN 61326-1:2006 |
| Safety compliance | ANSI/ISA 61010-1 / (82.02.01): 3rd edition CAN/CSA-C22.2 No 61010-1-12: 3rd edition UL 61010-1: 3rd edition IEC/EN 61010-1:2010 |
| Safety rating | CAT IV 600 V, CAT III 1000 V |
| Certifications | CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE |
| Ingress Protection (IP) rating | IP42 |
| Pollution Degree | 2 |
| Size (HxWxD) | 16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in) |
| Weight | .22 kg (8 oz) |
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| mV dc | 600.0 mV | 0.1 mV | 0.09 % + 3 |
| V dc | 6.000 V | 0.001 V | 0.09 % + 3 |
| 60.00 V | 0.01 V | ||
| 600.0 V | 0.1 V | ||
| 1000 V | 1 V | 0.15% + 2 |
| LCD w/backlight | 3.5 digits, 6000 counts, updates 4/sec |
|---|---|
| Battery type | 2 AA Alkaline batteries, NEDA 15A, IEC LR6 |
| Log rate/Interval | Adjustable by PC from 1 sec to 1 hr, default 1 min |
| Battery life | 400 Hours |
| Memory | Record up to 65000 readings |
| RF communications | 2.4 GHZ ISM Band |
| RF communication range | Open air, unobstructed: Up to 20 m Obstructed, sheetrock wall: Up to 6.5 m Obstructed, concrete wall or steel electrical enclosure: Up to 3.5 m |
| Operating temperature | -10 °C to +50 °C |
| Storage temperature | -40 °C to +60 °C |
| Temperature coefficient | 0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C) |
| Operating humidity | 90 % at 35 °C, 45 % at 40 °C, 45 % at 50 °C |
| EMC | IEC 61236-1, Portable |
| Safety compliance | IEC 61010-1, 600 V CAT IV / 1000 V CAT III, 3rd edition Pollution Degree 2 |
| Certifications | CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE |
| Ingress Protection (IP) rating | IP42 |
| Pollution Degree | 2 |
| Size (HxWxD) | 160 mm x 66 mm x 38 mm (6.3 in x 2.6 in x 1.5 in) |
| Weight | 0.255 kg (9 oz) |
For all Specifications accuracy is specified for one year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
| Accuracy 2,3,4 | |||
|---|---|---|---|
| Range1 | Resolution | 45 Hz to 500 Hz | 500 Hz to 1 kHz |
| 600.0 mV | 0.1 mV | 1.0 % + 3 | 2.0 % + 3 |
| 6.000 V | 0.001 V | ||
| 60.00 V | 0.01 V | ||
| 600.0 V | 0.1 V | ||
| 1000 V | 1 V | ||
1 All ac voltage ranges are specified from 1 % of range to 100 % of range.
2 Crest factor of < 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor up to 3.
4 Do not exceed 107 V-Hz.
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| mV | 600.0 mV | 0.1 mV | 0.09 % + 2 |
| V | 6.000 V | 0.001 V | 0.09 % + 2 |
| 60.00 V | 0.01 V | ||
| 600.0 V | 0.1 V | ||
| 1000 V | 1 V | 0.15 % + 2 | |
| Ω | 600 Ω | 1 Ω | Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 µs or longer. |
| Ω | 600.0 Ω | 0.1 Ω | 0.5 % + 2 |
| 6.000 kΩ | 0.001 kΩ | 0.5% + 1 | |
| 60.00 kΩ | 0.01 kΩ | ||
| 60.00 kΩ | 0.1 kΩ | ||
| 600.0 kΩ | 0.001 MΩ | ||
| 50.00 MΩ | 0.01 MΩ | 1.5 % + 3 | |
| Diode test | 2.000 V | 0.001 V | 1 % + 2 |
| µF | 1000 nF | 1 nF | 1.2 % + 2 |
| 10.00 µF | 0.01 µF | ||
| 100.0 µF | 0.1 µF | ||
| 9999 µF1 | 1 µF | 10 % typical |
1 In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % + 2.
| Function | Range1 | Resolution | Accuracy |
|---|---|---|---|
| mA AC (45 Hz to 1 kHZ) | 60.00 mA | 0.01 mA | 1.5 % + 3 |
| 400.0 mA3 | 0.1 mA | ||
| mA DC2 | 60.00 mA | 0.01 mA | 0.5 % + 3 |
| 400.0 mA3 | 0.1 mA |
1 All ac current ranges are specified from 5 % of range to 100 % of range.
2 Input burden voltage (typical): 400 mA input 2 mV/mA.
3 400.0 mA accuracy specified up to 600 mA overload.
| Range | Resolution | Accuracy [1] |
|---|---|---|
| 99.99 Hz | 0.01 Hz | 0.1 % + 1 |
| 999.9 Hz | 0.1 Hz | |
| 9.999 kHz | 0.001 kHz | |
| 99.99 kHz | 0.01 kHz |
1 Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.
| Function | Accuracy |
|---|---|
| DC functions | The specified accuracy of the measurement function ± 12 counts for changes > 350 ms in duration. |
| AC functions | The specified accuracy of the measurement function ± 40 counts for changes > 900 ms in duration. |
| Maximum voltage between any terminal and earth ground | 1000 V dc or ac rms |
|---|---|
| Ω fuse protection from A inputs | 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only |
| Display (LCD) | Update rate: 4/sec Votls, amps, ohms: 6000 counts Frequency: 10,000 counts Capacitance: 1,000 counts |
| Battery type | Three AA Alkaline batteries, NEDA 15A IEC LR6 |
| Battery life | 250 hours minimum |
| RF communications | 2.4 GHZ ISM Band |
| RF communication range | Open air, unobstructed: Up to 20m Obstructed, sheetrock wall: Up to 6.5m Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m |
| Temperature | Operating: -10 °C to 50 °C Storage: -40 °C to 60 °C |
| Temperature coefficient | 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
| Relative humidity | 0 % to 90 % (0 °C to 35 °C), 0 % to 75 % (35 °C to 40 °C), 0 % to 45 % (40 °C to 50 °C) |
| Altitude | Operating: 2,000 m Storage: 12,000 m |
| Electromagnetic compatibility EMI, RFI, EMC, RF | EN 61326-1:2006, EN 61326-2-2:2006. ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489 V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249, FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE |
| Safety compliance | ANSI/ISA 61010-1 / (82.02.01): 3rd edition CAN/CSA-C22.2 No 61010-1-12: 3rd edition UL 61010-1: 3rd edition IEC/EN 61010-1:2010 |
| Certifications | CSA, FCC, CE |
| Ingress Protection (IP) rating | IP54 |
| Pollution Degree | 2 |
| Size (H x W x L) | 4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in) |
| Weight | 340 g (12 oz) |


This website uses cookies (Learn more) and has a privacy policy (Learn more).