| Measurement Specifications | |
| FieldSense AC Voltage | Requires test leads: Yes, black lead Range: 1000 V Resolution: 1 V Accuracy: ±(3% + 3 counts), 45 to 66 Hz |
| AC Current | Requires test leads: No Range: 200 A Resolution: 0.1 A Accuracy: ±(3% + 3 counts), 45 to 66 Hz |
| FieldSense Frequency (Hz) | Requires test leads: No Range: 45 to 66 Hz Resolution: 1 Hz Accuracy: ±(1% + 2 counts) |
| Volts AC | Requires test leads: Yes Range: 1000 V Resolution: 1 V Accuracy: ±(1.5% + 2 counts), 45 to 66 Hz |
| Volts DC | Requires test leads: Yes Range: 1000 V Resolution: 1 V Accuracy: ±(1% + 2 counts) |
| Resistance | Requires test leads: Yes Range: 2000 Ω, 20 kΩ, 100 kΩ Resolution: 1 Ω, 0.01 kΩ, 0.1 kΩ Accuracy: ±(1% + 2 counts) |
| General Specifications | |
| Temperature | Operating: 14 to 122°F (-10 to 50°C) Storage: -22 to 140°F (-30 to 60°C) |
| Altitude | Operating: 6562' (2000 m) Storage: 32,808' (10,000 m) |
| Relative Humidity | 0 to 90%, 41 to 86°F (5 to 30°C) 0 to 75%, 86 to 104°F (30 to 40°C) 0 to 45%, 104 to 122°F (40 to 50°C) |
| Temperature Coefficient | 0.1 x (specified accuracy)/<64.4°F or >82.4°F (°C for <18°C or >28°C) |
| Safety | IEC/EN 61010-1, IEC/EN 61010-2-033: Pollution Degree 2 |
| Electromagnetic Environment | IEC/EN 61326-1: Portable |
| Battery | 2 x AA (IEC LR6); 360 hours continuous, typical; 200 hours using FieldSense |
| Fork Opening | 0.7" (17.8 mm) |
| Dimensions | 2.4 x 10.2 x 1.7" (61.1 x 259.3 x 43.8 mm) |
| Weight | 0.78 lb (0.35 kg) |