รายละเอียดสินค้า FLUKE 1745
|
Intrinsic error |
|
|
Refers to the reference conditions and is guaranteed for two years |
|
|
Warranty |
|
|
Recalibration interval |
|
|
Quality system |
|
|
developed, designed, and manufactured according to DIN ISO 9001 |
|
|
Reference conditions |
|
|
23 °C ± 2 K; 74 °F ± 2 K , Vm=230 V ± 10%, 50 Hz ± 0.1 Hz or 60 Hz ± 0.1 Hz |
|
Phase sequence: |
L1, L2, L3 |
|
Interval length: |
10 minutes |
|
Wye connection: |
(L1, L2, L3 to N) |
|
Power supply: |
88 V ... 265 V AC |
|
|
Environment conditions |
|
Working temp. range: |
-10 °C to 55°C; 14 °F to 131°F |
|
Operating temp. range: |
0 °C to 35°C; 32 °F to 95°F |
|
Storage temp. range: |
-20 °C to 60°C; -4 °F to 140°F |
|
Reference temp. range: |
23 °C ± 2 K; 74 °F ± 2 K |
|
Relative humidity: |
Fluke 1745: Class B2 acc. IEC 60654-1
Fluke 1744/43: Class C2 acc. IEC 60654-1 |
|
|
Housing |
|
|
robust, fully insulated housing and accessories |
|
|
Protection type |
|
Fluke 1745: |
IP50 as per EN 60529 |
|
Fluke 1744/1743: |
IP65 as per EN 60529 |
|
|
Safety |
|
|
IEC/EN 61010-1 600 V CAT III, 300 V CAT IV, pollution degree 2, double insulation |
|
|
Type test voltage |
|
|
5.2 kV rms, 50 Hz / 60 Hz, 5 s |
|
|
EMC |
|
Emission: |
IEC/EN 61326-1, EN55022 |
|
Immunity: |
IEC/EN 61326-1 |
|