Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
Includes variable volume for ἀne adjustment of pressures
Pressure measurement accuracy of 0.025 % reading ± 0.01 % FS
Electrical measurement and 24 volt supply for close looped calibrations
Measure or generate 4 to 20 mA
Measure 0 to 30 V dc
Powered by internal, rechargeable, high capacity NiMH battery or universal ac mains adapter
Compatible with Fluke 750P Pressure Modules